SRS DC205 Precision DC Voltage Source

Product Code:
64496
Manufacturer: Stanford Research Systems
  • ±100 VDC range
  • True 6-digit resolution
  • 1 ppm/°C stability
  • 0.0025 % accuracy (1 yr.)
  • Triggerable voltage scans
  • Low-noise design
  • Linear power supply
  • RS-232, USB and fiber optic interfaces

More info and finance questions

Request a Quote

Lambda Exclusive Promotion:
Additional 12 months warranty for free (2 years total) via our UK Service Centre.

The SRS DC205 low-noise, high-resolution DC voltage source is the right tool when a precision bias source is needed. Its bipolar, four-quadrant output delivers up to 100 V with 1 µV resolution and up to 50 mA of current. In 4-wire mode (remote sense), the instrument corrects for lead resistance delivering accurate potential to your load. The DC205's output stability is a remarkable ±1 ppm over 24 hours. With its linear power supply, there is no need to worry about high-frequency noise.

True 6-Digit Resolution
The front-panel display of the DC205 lets you set voltage with true 6-digit resolution. There are three voltage ranges to choose from: ±1 V, ±10 V and ±100 V which allows voltage settings from 1 µV to 100 V — eight orders of magnitude!

Low-Noise Design
The DC205 has outstanding noise characteristics - on the 1 V range, the rms noise is less than 1 µV (0.1 Hz to 10 Hz). It is also accurate to 0.0025 % over a one year period, and it has excellent temperature stability with a specification of less than 1 ppm/°C. The design even features linear power supplies rather than switching power supplies, so switching frequency interference can never be a problem.

Scanning
The instrument's triggerable voltage scanning feature can be useful in a number of experimental applications. The start and stop voltage, scan speed, and scan function can all be controlled. Scan speeds can be set from ten seconds to thirteen hours, and the scan function can either be a ramp or a triangle wave. Single scans and continuous scans are both supported, and the instrument can be triggered from the front panel, remotely over one of the interfaces, or from an external trigger signal.

Bipolar, Four-Quadrant Output
The DC205 can output either positive or negative voltages, and it operates in either grounded or floating mode. In floating mode, the output can float up to 250 V relative to chassis ground. You can also select either 2-wire or 4-wire operation. In 4-wire mode (remote sense), the instrument maintains its preset voltage directly at your load eliminating the effect of lead resistance.

Computer Interfaces
The DC205 has both RS-232 and USB computer interfaces on its rear panel. All functions of the instrument can be set or read via the interfaces. For remote interfacing with complete electrical isolation, the DC205 also has a rear-panel fiber optic interface. When connected to the SX199 Remote Computer Interface Unit, a path for controlling the DC205 via GPIB, Ethernet, and RS-232 is provided.

Signal Output

Output

2-wire or 4-wire (remote sense)
Output can be set to Ground or Float (250 volts max.) mode

±1 V Range

Full scale

±1.010000 V

Resolution

1 µV

Current

50 mA (max.)

Accuracy

24 hour: ±(7 ppm of setting + 2 µV)
90 day*: ±(12 ppm of setting + 6 µV)
1 year*: ±(25 ppm of setting + 10 µV)

Stability

24 hour: ±(1 ppm of setting + 1 µV)

Temp. Co.

±(1 ppm of setting + 1 µV) / °C
(0 °C to 40 °C)

Noise (typ.)

0.5 µVrms (0.1 Hz to 10 Hz)
9 µVrms (10 Hz to 100 kHz)

±10 V Range

Full scale

±10.10000 V

Resolution

10 µV

Current

50 mA (max.)

Accuracy

24 hour: ±(7 ppm of setting + 6 µV)
90 day*: ±(12 ppm of setting + 15 µV)
1 year*: ±(25 ppm of setting + 20 µV)

Stability

24 hour: ±(1 ppm of setting + 3 µV)

Temp. Co.

±(1 ppm of setting + 2 µV) / °C
(0 °C to 40 °C)

Noise (typ.)

1.2 µVrms (0.1 Hz to 10 Hz)
12 µVrms (10 Hz to 100 kHz)

±100 V Range

Full scale

±101.0000 V

Resolution

100 µV

Current

25 mA (max.)

Accuracy

24 hour: ±(8 ppm of setting + 60 µV
90 day*: ±(12 ppm of setting + 80 µV)
1 year*: ±(25 ppm of setting + 100 µV)

Stability

24 hour: ±(1 ppm of setting + 20 µV)

Temp. Co.

±(1 ppm of setting + 15 µV) / °C
(0 °C to 40 °C)

Noise (typ.)

12 µVrms (0.1 Hz to 10 Hz)
50 µVrms (10 Hz to 100 kHz)

Voltage Scanning

Scan rate

10 s to 13 hrs.

Scan type

Ramp or triangle wave, continuous or single shot

Triggering

Scans can be triggered using the rear-panel trigger input, from the front panel, or over one of the computer interfaces

Remote Interfaces

USB

Virtual COM port with FTDI drivers, 115.2k baud, 8 bits, no parity, 1 stop bit, RTS/CTS flow

RS-232

DB-9 connector, 9600 baud

Fiber Optic

Connection to SX199 Optical Interface Controller. Provides isolated connectivity to GPIB, RS-232 and Ethernet

General

Operating temp.

0 °C to 40 °C, non-condensing

Power

<30 W, 100/120/220/240 VAC, 50 Hz or 60 Hz

Dimensions

8.3" × 3.5" × 13.0" (WHL)

Weight

10 lbs.

Warranty

One year parts and labor on defects in materials and workmanship

* Preliminary specifications

Write Your Own Review